Dosimetric and qualitative analysis of kinetic properties of millennium 80 multileaf collimator system for dynamic intensity modulated radiotherapy treatments
Anup K Bhardwaj1, TS Kehwar2, SK Chakarvarti3, AS Oinam1, SC Sharma1
1 Department of Radiotherapy, Post-graduate Institute of Medical Education and Research, Chandigarh -160 012, India 2 Department of Radiation Oncology, University of Pittsburgh Cancer Institute, Pittsburgh, Pennsylvania, USA 3 Department of Applied Physics, National Institute of Technology, Kurukshetra - 136 119, India
Correspondence Address:
Anup K Bhardwaj Medical Physicist and RSO, Department of Radiotherapy, Post Graduate of Institute of Medical Education and Research, Chandigarh - 160012 India
 Source of Support: None, Conflict of Interest: None  | Check |
DOI: 10.4103/0973-1482.31967
|
|
The aim of this paper is to analyze the positional accuracy, kinetic properties of the dynamic multileaf collimator (MLC) and dosimetric evaluation of fractional dose delivery for the intensity modulated radiotherapy (IMRT) for step and shoot and sliding window (dynamic) techniques of Varian multileaf collimator millennium 80. Various quality assurance tests such as accuracy in leaf positioning and speed, stability of dynamic MLC output, inter and intra leaf transmission, dosimetric leaf separation and multiple carriage field verification were performed. Evaluation of standard field patterns as pyramid, peaks, wedge, chair, garden fence test, picket fence test and sweeping gap output was done. Patient dose quality assurance procedure consists of an absolute dose measurement for all fields at 5 cm depth on solid water phantom using 0.6cc water proof ion chamber and relative dose verification using Kodak EDR-2 films for all treatment fields along transverse and coronal direction using IMRT phantom. The relative dose verification was performed using Omni Pro IMRT film verification software. The tests performed showed acceptable results for commissioning the millennium 80 MLC and Clinac DHX for dynamic and step and shoot IMRT treatments. |
|
|
|
[FULL TEXT] [PDF]* |
|
 |
|