Analysis of induced error by susceptibility effect in low-density gel dosimeters


1 Department of Radiation Science, The School of Allied Medical Sciences, Tehran University of Medical Sciences, Tehran, Iran
2 Department of Radiotherapy Oncology, Cancer Research Centre, Cancer Institute, Tehran University of Medical Sciences, Tehran, Iran
3 Department of Medical Physics and Biomedical Engineering, Faculty of Medicine, Tehran University of Medical Sciences, Tehran, Iran; WAVES Research Group, Department of Information Technology, Ghent University, Ghent, Belgium
4 Department of Food Science and the Rutgers Center for Lipid Research, New Jersey Institute for Food, Nutrition, and Health, Rutgers University, New Brunswick, New Jersey 08901, USA

Correspondence Address:
Farideh Pak,
Department of Radiation Science, The School of Allied Medical Sciences, Tehran University of Medical Sciences, West Taleghani Avenue, Poursina Street, Tehran 14174
Iran
Login to access the Email id

Source of Support: None, Conflict of Interest: None

DOI: 10.4103/jcrt.JCRT_407_17

 > Abstract 


Purpose: In low-density (LD) gel dosimeter, diffusive spin–spin relaxation rate (R2)-dispersion caused by susceptibility-induced internal gradient leads to a significant deviation in the measured R2 from the real value. In this study, the effect of induced internal gradient on R2 was visualized and quantified algebraically as an important cause of inaccuracy in LD gel dosimeters.
Materials and Methods: In this method, two sets of LD and unit-density (UD) gel dosimeters were prepared. The LD gel was made by mixing the UD gel with expanded polystyrene spheres. The R2 was used to determine the spatially resolved decay rates due to diffusion in internal magnetic field. The internal gradient was calculated for a multiple spin–echo sequence.
Results: It is shown that in a LD gel, the internal gradient leads to overestimation of mean R2 value (R2mean). Pixel-by-pixel R2 measurements inside a LD gel showed significant deviation from R2 mapping in UD gel.
Conclusion: It appears that significant differences between R2meanin a selected region of interest and pixel-by-pixel R2 values are the main source of inaccuracy in dose mapping of a LD gel.

Keywords: Dose mapping, inaccuracy, low-density gel dosimeter, mathematical analysis, susceptibility effect



How to cite this URL:
Pak F, Vaezzadeh V, Eqlimi E, Mirheydari M. Analysis of induced error by susceptibility effect in low-density gel dosimeters. J Can Res Ther [Epub ahead of print] [cited 2019 Mar 21]. Available from: http://www.cancerjournal.net/preprintarticle.asp?id=244214

.






 

 
Top
 
 
  Search
 
     Search Pubmed for
 
    -  Pak F
    -  Vaezzadeh V
    -  Eqlimi E
    -  Mirheydari M
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

  >Abstract
  In this article

 Article Access Statistics
    Viewed234    
    PDF Downloaded2    

Recommend this journal