Commissioning of portal dosimetry using a novel method for flattening filter-free photon beam in a nontrue beam linear accelerator


1 Department of Radiation Oncology, Yashoda Hospitals, Hyderabad, Telangana; Research and Development Centre, Bharathiar University, Coimbatore, Tamil Nadu, India
2 Research and Development Centre, Bharathiar University, Coimbatore, Tamil Nadu; Department of Radiotherapy, All India Institute of Medical Sciences, New Delhi, India
3 Department of Radiation Oncology, Yashoda Hospitals, Hyderabad, Telangana, India

Correspondence Address:
Subramani Vellaiyan,
All India Institute of Medical Sciences, Ansari Nagar East, New Delhi - 110 029
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None

DOI: 10.4103/jcrt.JCRT_1181_16

 > Abstract 



Aim: The aim of this study is to commission and validate the portal dosimetry (PD) system using an indirect method for flattening filter free (FFF) photon beam of the upgraded c-series linear accelerator.
Background: Varian Medical System clinacs with amorphous-silicon portal imager panel (aSi-1000) do not have PD for FFF beams. Recently, our c-series linear accelerator was upgraded to deliver 6MV FFF (6MVFFF) photon beam with the highest dose rate of 1400 monitor unit (MU)/min. The study, therefore, focuses on the commissioning and validation of PD for the 6MVFFF beam.
Materials and Methods: An indirect method was implemented to predict the portal dose for FFF beam in Eclipse as the treatment planning system does not have direct prediction algorithm for FFF beam (version. 11). Dosimetrical characteristics of aSi-electronic portal imaging device (EPID) were evaluated for 6MVFFF beam and validation of PD for 6MVFFF beam was performed for open fields along with pretreatment quality assurance of intensity-modulated radiation therapy (IMRT), volumetric-modulated arc therapy (VMAT), and stereotactic radiosurgery (SRS) techniques for 30 patients planned with 6MVFFF beam.
Results:ASi-EPID saturates between 100 and 130 cm source to detector distance (SDD) for 6MVFFF beam and resolved at more than 140 cm SDD. The squared correlation coefficient (R2) for MU linearity was found to be 1 (R2 = 1), and instantaneous dose response linearity at different SDD's was found to be 0.999 (R2 = 0.999) for the 6MVFFF beam. Maximum gamma area index (GAI) for 3% dose difference and 3 mm distance-to-agreement criteria for IMRT, VMAT, and SRS/stereotactic radiotherapy plans was 97.9% ± 0.3%, 96.3% ± 0.5%, and 98.2% ± 0.2%, respectively.
Conclusion: The results reveal that this novel method can be used to commission portal dosimetry for 6MVFFF beam as it is a convenient, faster, and accurate method.

Keywords: ASi-electronic portal imaging device, flattening filter free beam, portal dosimetry, volumetric-modulated arc therapy



How to cite this URL:
Gandhi A, Vellaiyan S, Subramanian V S, Shanmugam T, Murugesan K, Kala S. Commissioning of portal dosimetry using a novel method for flattening filter-free photon beam in a nontrue beam linear accelerator. J Can Res Ther [Epub ahead of print] [cited 2019 Jan 20]. Available from: http://www.cancerjournal.net/preprintarticle.asp?id=241956

.




 

 
Top
 
 
  Search
 
     Search Pubmed for
 
    -  Gandhi A
    -  Vellaiyan S
    -  Subramanian V S
    -  Shanmugam T
    -  Murugesan K
    -  Kala S
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

  >Abstract
  In this article

 Article Access Statistics
    Viewed224    
    PDF Downloaded12    

Recommend this journal