ORIGINAL ARTICLE
Year : 2019  |  Volume : 15  |  Issue : 5  |  Page : 1011-1017

Two-dimensional dose reconstruction using scatter correction of portal images


1 Department of Radiation Oncology, University of Maryland School of Medicine, Baltimore, MD, USA
2 Department of Radiation Oncology, University of Minnesota, Minneapolis, MN, USA
3 Department of Medical Physics, Royal Adelaide Hospital, Adelaide, SA, Australia

Correspondence Address:
Akbar Anvari
Department of Radiation Oncology, University of Maryland School of Medicine, Baltimore, MD
USA
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/jcrt.JCRT_376_17

Rights and Permissions

Context: Electronic portal imaging devices (EPIDs) could potentially be useful for patient setup verification and are also increasingly used for dosimetric verification. The accuracy of EPID for dose verification is dependent on the dose-response characteristics, and without a comprehensive evaluation of dose-response characteristics, EPIDs should not be used clinically. Aims: A scatter correction method is presented which is based on experimental data of a two-dimensional (2D) ion chamber array. An accurate algorithm for 2D dose reconstruction at midplane using portal images for in vivo dose verification has been developed. Subjects and Methods: The procedure of scatter correction and dose reconstruction was based on the application of several corrections for beam attenuation, and off-axis factors, measured using a 2D ion chamber array. 2D dose was reconstructed in slab phantom, OCTAVIUS 4D system, and patient, by back projection of transit dose map at EPID-sensitive layer using percentage depth dose data and inverse square. Verification of the developed algorithm was performed by comparing dose values reconstructed in OCTAVIUS 4D system and with that provided by a treatment planning system. Results: The gamma analysis for dose planes within the OCTAVIUS 4D system showed 98% ±1% passing rate, using a 3%/3 mm pass criteria. Applying the algorithm for dose reconstruction in patient pelvic plans showed gamma passing rate of 96% ±2% using the same pass criteria. Conclusions: An accurate empirical algorithm for 2D patient dose reconstruction has been developed. The algorithm was applied to phantom and patient data sets and is able to calculate dose in the midplane. Results indicate that the EPID dose reconstruction algorithm presented in this work is suitable for clinical implementation.


[FULL TEXT] [PDF]*
Print this article     Email this article
 Next article
 Previous article
 Table of Contents

 Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
 Citation Manager
 Access Statistics
 Reader Comments
 Email Alert *
 Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed1027    
    Printed42    
    Emailed0    
    PDF Downloaded30    
    Comments [Add]    

Recommend this journal