Evaluation of on-board imager cone beam CT hounsfield units for treatment planning using rigid image registration


Department of Radiation Therapy, Christian Medical College, Vellore, Tamil Nadu, India

Correspondence Address:
Mohamathu Rafic
Department of Radiation Therapy, Christian Medical College, Vellore - 632 004, Tamil Nadu
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0973-1482.146087

Rights and Permissions

Purpose: To evaluate the on-board imager cone beam CT (OBI-CBCT) Hounsfield units (HUs) for treatment planning. Materials and Methods: The HU-electron density (eD) calibration for CBCT, the CATphan504 phantom was used, and the CBCT HU (HU CBCT ) consistency was studied by analyzing the CBCT images of Rando phantom and compared with planning CT. The latter study was also performed on CBCT images of 10 H&N patients. For comparison, the structures contoured and treatment plans generated on CT were transferred on to the CBCT after registration. The treatment plans were compared using gamma (g) index analysis and the plan comparison dose volume histograms (DVH PlanComp ). Results: Although the HU-eD calibration curves of both the planning CT and CBCT were found to be linear, differences in mean HU values were found in the region of interest (ROI) corresponding to Acrylic, Derlin, and Teflon, viz., 144 ± 11 HU, 193 ± 5 HU, and 257 ± 7 HU respectively. For all the cases, the consistency and reproducibility of HU CBCT values for low density medium agreed the HU CT except at regions of high density. Overall g-evaluation showed more than 94% pixels pass rate and DVH results showed small difference in the DVH PlanComp, Rando, and large differences in DVH PlanComp, patient for structures contoured at peripheral regions (PV) of CBCT images. Conclusions: We conclude that the pixel-to-pixel HU corrections for entire range of eD are not necessary for OBI-CBCT images. Application of local correction in the high-density and penumbral regions would facilitate the use of CBCT images for routine treatment planning.


[FULL TEXT] [PDF]*
Print this article     Email this article
 Next article
 Previous article
 Table of Contents

 Similar in PUBMED
 Related articles
 Citation Manager
 Access Statistics
 Reader Comments
 Email Alert *
 Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed3858    
    Printed63    
    Emailed0    
    PDF Downloaded211    
    Comments [Add]    

Recommend this journal