Does the fluence map editing in electronic tissue compensator improve dose homogeneity in bilateral field plan of head and neck patients?


Department of Medical Physics, Tata Memorial Hospital, Parel, Mumbai, India

Correspondence Address:
Rajesh A Kinhikar
Department of Medical Physics, Tata Memorial Hospital, Parel, Mumbai - 400 012
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0973-1482.42254

Rights and Permissions

The purpose of this study was to evaluate the effect of fluence map editing in electronic tissue compensator (ETC) on the dose homogeneity for head and neck cancer patients. Treatment planning using 6-MV X-rays and bilateral field arrangement employing ETC was carried out on the computed tomography (CT) datasets of 20 patients with head and neck cancer. All the patients were planned in Varian Eclipse three-dimensional treatment planning system (3DTPS) with dynamic multileaf collimator (DMLC). The treatment plans, with and without fluence editing, was compared and the effect of pre-editing and post-editing the fluence maps in the treatment field was evaluated. The skin dose was measured with thermoluminescent dosimeters (TLDs) and was compared with the skin dose estimated by TPS. The mean percentage volume of the tissue receiving at least 107% of the prescription dose was 5.4 (range 1.5-10; SD 2.4). Post-editing fluence map showed that the mean percentage volume of the tissue receiving at least 107% of the prescription dose was 0.47 (range 0.1-0.9; SD 0.3). The mean skin dose measured with TLD was found to be 74% (range 71-80%) of the prescribed dose while the TPS showed the mean skin dose as 85% (range 80-90%). The TPS overestimated the skin dose by 11%. Fluence map editing thus proved to be a potential tool for improving dose homogeneity in head and neck cancer patients planned with ETC, thus reducing the hot spots in the treatment region as well. The treatment with ETC is feasible with DMLC and does not take any additional time for setup or delivery. The method used to edit the fluence maps is simple and time efficient. Manual control over a plan is essential to create the best treatment plan possible.


[FULL TEXT] [PDF]*
Print this article     Email this article
 Next article
 Previous article
 Table of Contents

 Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
 Citation Manager
 Access Statistics
 Reader Comments
 Email Alert *
 Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed3968    
    Printed144    
    Emailed1    
    PDF Downloaded262    
    Comments [Add]    

Recommend this journal